Troubleshooting
11 Troubleshooting
This section on troubleshooting is a guide to problems associated with the installation and integration of
the system only. Section 4 of the PH10 series user’s guide (Renishaw part number H-1000-5070) deals
with problems associated with normal operation of the system.
The optional HCU1 can also be used to identify system faults. For full details of the use and fault finding
capability of the unit see the HCU1 hand control unit user’s guide (Renishaw part number H-1000-5016).
Use table 25 to identify problems you are experiencing with your system. If you experience problems
which you are not able to identify or solve satisfactorily, please contact Renishaw for further advice.
Table 25 - Troubleshooting
Observation Possible cause Checks / remedies
POWER ON LED
not lit.
Power loss. Check mains cable connections and integrity.
Check PHC10-2 fuse (see page 6).
Check CMM EMERGENCY STOP condition - power may
have been removed by the CMM.
No head
movement in
automatic mode.
Power loss. See above.
Cable/connection fault. Check connections and integrity of cabling between head
and PHC10-2.
Communications failure. Check RS232 baud rate (see section 4.2) or IEEE protocol
selection for RS232 (see section 5.3).
Incorrect command set
selected.
Check correct command set is selected (see section 4.3.1).
No head
movement in
manual mode.
Power loss. See above.
Cable/connection fault. Check connections and integrity of cabling between head
and PHC10-2.
No output signal
received by the
CMM computer.
Cable/connection fault. Check connections and integrity of cabling from:
- head to PHC10-2
- PHC10-2 to interface
- interface to CMM computer
Multiwire bypass
connector not fitted.
When using standard touch-trigger probes, the probe head
multiwire bypass connector must be fitted to permit the
probe signal to reach the PHC10-2.
Multiwired probe in use. Check that the multiwire cable is correctly fitted to the head.
Check that the trigger output to the CMM computer is
connected to the multiwired probe interface.
Poor
measurement
performance.
Loose mounting of
head.
Ensure all mounting screws are tight and mounting to CMM
is secure.
Probe DAMPing
enabled during
measurement.
Ensure probe DAMPing is not enabled during
measurement moves.