17
RS232 serial communications
4.3.1 Basic command set mode
In this mode the PHC10-2 is fully compatible with existing integration methods in terms of
communications protocols, software command set and RS232 protocol options.
To select the basic command set mode and protocol, set switch 9 (command set selection) DOWN.
When this mode is chosen, the following protocol options are available (table 4).
Table 4 - Basic command set protocol
Switch Position Selection
6
UP 2 stop bits
DOWN 1 stop bit
7
UP CTS (Clear To Send) ON
DOWN CTS (Clear To Send) OFF
8
UP LF (Line Feed) ON
DOWN LF (Line Feed) OFF
NOTE: Switches 6, 7 and 8 have no effect when switch 9 is UP.
Data transmission format is as follows:
1 start bit
7 data bits
1 parity bit (ignored on Rx: always 0 on Tx)
1 or 2 stop bits (switch 6)
Protocols:
PHC10-2 RTS is asserted before first transmission from the head and remains asserted. If switch 7
is UP, CTS must be asserted by the CMM computer to allow the head to transmit. If switch 8 is UP,
the PHC10-2 will add an ASCII LF character to every transmitted message.
4.3.2 Extended command set mode
In this mode the PHC10-2 uses the Renishaw extended command set. It is completely different from and
incompatible with the basic command set. The RS232 protocol is fixed with no user selectable options
other than baud rate (see section 4.2).
NOTE: The extended command set has not been implemented on the IEEE version of the PHC10-2.